DEFIBRILLATOR Waveform: Biphasic: Low current, impedance compensated, optimized
terminating current, truncated exponential biphasic Energy Select: External: 2, 5, 7, 10, 20, 30, 50, 70, 100, 150,
200, 300, 360 J. Internal: 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50 J. Charge Time: < 10 seconds @ 360 J (typical) Charge
Indicator: Audible and graphic. Output: Adult paddles, internal paddles, peds adapter and multipurpose hands-free adapter
available. Synchronizer: Delivers energy within 60 msec. Disarm: Front panel switch. MONITOR/DISPLAY Input:
3-lead PIC: 3-lead or 5-lead patient cable, paddles, or multipurpose hands-free adapter. 12-lead PIC: 3-lead, 12-Lead
patient cable, paddles or mutipurpose hands free adapter. Size: 6.5inch; 16.5cm diagonal, non-fade. Type: Color active
matrix TFT display Display Resolution: 640 x 480 pixels. Sweep Speed: 25 mm / sec. Lead Selections: Paddles (Pads),
I, II, III, AVR, AVL, AVF, V. Frequency Response: (User-selectable). • 2 to 20 Hz Limited mode • 1 to
40 Hz Monitor mode • 0.05 to 40 Hz Limited Diagnostic mode • 0.05 to 150 Hz Diagnostic mode (automatically
sets chart recorder response) Common Mode Rejection: 90dB (Complies with AAMI EC13-1992 section 3.2.9.10.) Heart
Rate Meter: 20 to 300 BPM. Heart Rate Alarms: User-selectable. Size: 0.125, 0.25, 0.5, 1, 2, 4 cm/mv and auto-sizing. ECG
Output: 1 V/mV RECORDER Type: High-resolution thermal array. Annotation: Time, date, ECG lead, ECG gain, heart
rate, defibrillation and pacing parameters and treatment summary ACLS events. Paper Width: 50 mm. Paper Speed: 25
mm/sec. Delay: 6 seconds. Frequency Response: Automatically set to monitor's frequency response. Treatment Summary:
7 switches to record key ACLS events (IV, INTUB, EPI, LIDO, ATROP, etc.). Automatically logs into memory the type of
event, time and ECG sample. Tx Summary Log: 60 ECG events or 4000 non-ECG events. BATTERY High Capacity
Type: NiMH 12 volt - Welch Allyn SuperPac™. Low Battery Indicator: Flashing low battery icon on display. Multiple
Paired Contacts: Ensures quick, error-free insertion and backup reliability. Recharge Rate: 80% in 7.5 hours, 100% in
9.5 hours. (For high capacity batteries) GENERAL Weight: 10 lbs; 4.5 kg (basic configuration). Size: 12.5
x 12.5 x 5.3 inches; 32 x 32 x 13 cm Operating: 2.0 hours of continous ECG, SPO2, CO2, and BP monitoring or 120 shocks
at 360Joules (typical) Actual operating times will vary according to the condition of the battery and the number of
additional options activated. Temperature*: 0 to 45° C Humidity* (NC): 15 to 95% RH Vibration*: MIL-STD 810E Shock
/ Drop*: MIL-STD 810E Altitude: Up to 15,000 feet Transport and Storage: Temperature (-30 to 70˚C), humidity
(15 to 95% non-condensing), atmospheric pressure ( 860 to 1060 hPa), shock/vibration (ISTA 1A) Enclosure Protection*: Solid
Foreign Object: IEC 529, IP2X Water: IEC 529, IPX4. * All tests performed per AAMI DF-2 Defibrillation Standard. PULSE
OXIMETER Measurement Range: 31 to 99% Heart Rate Range: 30 to 250 BPM ±5% or 5 BPM whichever is less. Accuracy:
81-99% = ±2%, 70-80% = 3%, 0-69% unspecified. Update Period: Every pulse after 4 valid pulses. Alarm Limits: 60 lower
limit, 70 upper limit to 99%.
|