Monitech, Inc.
PIC50 Manufacturer's Specifications
Home | Monitor/Defibrillator Brochures | Monitech Staff | Products List | Used Equip | Territories | Contact Us

 

Technical Specifications

DEFIBRILLATOR
Waveform:
Biphasic: Low current, impedance compensated,
optimized terminating current, truncated exponential
biphasic
Energy Select:
External: 2, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200,
300, 360 J.
Internal: 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50 J.
Charge Time: < 10 seconds @ 360 J (typical)
Charge Indicator: Audible and graphic.
Output: Adult paddles, internal paddles, peds adapter
and multipurpose hands-free adapter available.
Synchronizer: Delivers energy within 60 msec.
Disarm: Front panel switch.
MONITOR/DISPLAY
Input: 3-lead PIC: 3-lead or 5-lead patient cable,
paddles, or multipurpose hands-free adapter.
12-lead PIC: 3-lead, 12-Lead patient cable, paddles or
mutipurpose hands free adapter.
Size: 6.5inch; 16.5cm diagonal, non-fade.
Type: Color active matrix TFT display
Display Resolution: 640 x 480 pixels.
Sweep Speed: 25 mm / sec.
Lead Selections: Paddles (Pads), I, II, III, AVR, AVL, AVF, V.
Frequency Response: (User-selectable).
• 2 to 20 Hz Limited mode
• 1 to 40 Hz Monitor mode
• 0.05 to 40 Hz Limited Diagnostic mode
• 0.05 to 150 Hz Diagnostic mode (automatically sets chart recorder response)
Common Mode Rejection: 90dB (Complies with AAMI
EC13-1992 section 3.2.9.10.)
Heart Rate Meter: 20 to 300 BPM.
Heart Rate Alarms: User-selectable.
Size: 0.125, 0.25, 0.5, 1, 2, 4 cm/mv and auto-sizing.
ECG Output: 1 V/mV
RECORDER
Type: High-resolution thermal array.
Annotation: Time, date, ECG lead, ECG gain, heart rate,
defibrillation and pacing parameters and treatment
summary ACLS events.
Paper Width: 50 mm. Paper Speed: 25 mm/sec.
Delay: 6 seconds.
Frequency Response: Automatically set to monitor's
frequency response.
Treatment Summary: 7 switches to record key ACLS
events (IV, INTUB, EPI, LIDO, ATROP, etc.).
Automatically logs into memory the type of event, time
and ECG sample.
Tx Summary Log: 60 ECG events or 4000 non-ECG events.
BATTERY
High Capacity Type: NiMH 12 volt - Welch Allyn
SuperPac™.
Low Battery Indicator: Flashing low battery icon
on display.
Multiple Paired Contacts: Ensures quick, error-free
insertion and backup reliability.
Recharge Rate: 80% in 7.5 hours, 100% in 9.5 hours.
(For high capacity batteries)
GENERAL
Weight: 10 lbs; 4.5 kg (basic configuration).
Size: 12.5 x 12.5 x 5.3 inches; 32 x 32 x 13 cm
Operating: 2.0 hours of continous ECG, SPO2, CO2, and
BP monitoring or 120 shocks at 360Joules (typical)
Actual operating times will vary according to the
condition of the battery and the number of additional
options activated.
Temperature*: 0 to 45° C
Humidity* (NC): 15 to 95% RH
Vibration*: MIL-STD 810E
Shock / Drop*: MIL-STD 810E
Altitude: Up to 15,000 feet
Transport and Storage: Temperature (-30 to 70˚C),
humidity (15 to 95% non-condensing), atmospheric
pressure ( 860 to 1060 hPa), shock/vibration (ISTA 1A)
Enclosure Protection*:
Solid Foreign Object: IEC 529, IP2X
Water: IEC 529, IPX4.
* All tests performed per AAMI DF-2 Defibrillation Standard.
PULSE OXIMETER
Measurement Range: 31 to 99%
Heart Rate Range: 30 to 250 BPM ±5% or 5 BPM whichever is less.
Accuracy: 81-99% = ±2%, 70-80% = 3%, 0-69% unspecified.
Update Period: Every pulse after 4 valid pulses.
Alarm Limits: 60 lower limit, 70 upper limit to 99%.

pic50.jpg

If you would like to receive more information on any of the products featured on our website, please click here.

Software * Services * Technology for the Medical Community