GENERAL Weight: 10.5 lbs (4.76 kg) basic configuration. Size:
16.7 x 12.5 x 5.3 inches; 41.75 x 32 x 13 cm. Operating: Time: 2 hours Temperature*: 0 to 45° C Humidity* (NC):
15 to 95% RH Vibration*: MIL-STD 810E Shock / Drop*: MIL-STD 810E Altitude: Up to 15,000 feet Transport and Storage:
Temperature (-30 to 70˚C), humidity (15 to 95% non-condensing), atmospheric pressure ( 860 to 1060 hPa), shock/vibration
(ISTA 1A) Enclosure Protection*: Solid Foreign Object: IEC 529, IP2X Water: IEC 529, IPX4. MONITOR/DISPLAY Input:
3-lead or 5-lead patient cable, paddles, or multipurpose hands-free adapter. Size: 6.4 inch (16.5 cm) diagonal, non-fade. Type:
LCD color active matrix. Display Resolution: 640 x 480 pixels. Sweep Speed: 25 mm / sec. Lead Selections: Paddles
(Pads), I, II, III, AVR, AVL, AVF, V. Frequency Response: (User-selectable). • 2 to 20 Hz Limited mode •
0.5 to 40 Hz Monitor mode • 0.05 to 150 Hz Diagnostic mode(automatically sets chart recorder response) Common
Mode Rejection: Complies with AAMI EC13- 1992 section 3.2.9.10. Heart Rate Meter: 20 to 300 BPM. Heart Rate Alarms:
User-selectable. Size: 0.125, 0.25, 0.5, 1, 2, 4 cm/mv and auto-ranging. AED Configurable: Basic & Basic+
modes Operation: Continuous analysis of ECG advises shock on VFIB, shockable VTACH Energy Sequence: User configurable Biphasic:
150/200/300/360J Factory Default: Biphasic: 150/200/300J Prompts: Audible voice and visual text prompts DEFIBRILLATOR Waveform: Biphasic:
Low current, impedance compensated, optimized terminating current, truncated exponential biphasic Energy Select: External:
2, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, 360 J. Internal: 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50 J. Charge
Time: < 10 seconds @ 360 J (typical) SuperPak Output: Adult paddles, internal paddles, peds adapter and multipurpose
hands-free adapter available. Synchronizer: Delivers energy within 60 msec. Disarm: Front panel switch. RECORDER Type:
High-resolution thermal array. Annotation: Time, date, ECG lead, ECG gain, heart rate, defibrillation and pacing parameters
and treatment summary ACLS events. Paper Width: 50 mm. Paper Speed: 25 mm/sec. Delay: 6 seconds. Frequency
Response: Automatically set to monitor's frequency response. Tx Summary Log: 60 ECG events or 300 non-ECG events. BATTERY High
Capacity Type: NiMH 12 volt - Welch Allyn SuperPac™. Low Battery Indicator: Flashing low battery icon on display. Multiple
Paired Contacts: Ensures quick, error-free insertion and backup reliability. Recharge Rate: 80% in 7.5 hours, 100% in
9.5 hours. (For high capacity batteries) PULSE OXIMETER Measurement Range: 31 to 99%. Heart Rate Range:
30 to 250 BPM ±5% or 5 BPM whichever is less. Accuracy: 81-99% = ±2%, 70-80% = 3%, 0-69% unspecified Update Period:
Every pulse after 4 valid pulses Alarm Limits: 60 lower limit, 70 upper limit to 99% Probe: Reusable finger or multi-site
probe
|
 |
|