Monitech, Inc.
PIC30 Manufacturer's Specifications
Home
Monitor/Defibrillator Brochures
Monitech Staff
Products List
Used Equip
Territories
Contact Us

 

Technical Specifications

GENERAL
Weight: 13.5 lbs (6.1 kg) (basic configuration).
Size: 16.7 x 12.5 x 5.3 inches; 41.75 x 32 x 13 cm.
Operating:
Time: 2 hours
Temperature*: 0 to 45° C
Humidity* (NC): 15 to 95% RH
Vibration*: MIL-STD 810E
Shock / Drop*: MIL-STD 810E
Altitude: Up to 15,000 feet
Transport and Storage: Temperature (-30 to 70˚C), humidity
(15 to 95% non-condensing), atmospheric pressure ( 860 to 1060 hPa), shock/vibration (ISTA 1A)
Enclosure Protection*:
Solid Foreign Object: IEC 529, IP2X
Water: IEC 529, IPX4.
* All tests performed per AAMI DF-2 Defibrillation Standard.
Operating Power: Battery pack or auxiliary power supply.

DEFIBRILLATOR
Waveform:
Biphasic: Low current, impedance compensated, optimized
terminating current, truncated exponential biphasic.
Energy Select:
External: 2, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, 360 J.
Internal: 2, 3, 4, 5, 6, 7, 8, 9, 10, 20, 30, 50 J.
Charge Time: < 7 seconds @ 360 J (with a charged battery).
< 15 seconds @ 360 J (no battery, 90% AC mains voltage)
Charge Indicator: Audible and graphic.
Output: Adult paddles, internal paddles, peds adapter and multipurpose hands-free
adapter available.
Synchronizer: Delivers energy within 60 msec.
Disarm: Front panel switch.

POWER
Mains Voltage: 100-240VAC (auto ranging)
Mains Frequency: 50/60Hz

BATTERY - (User replaceable)
High Capacity Type: NiMH 12 volt - Welch Allyn
SuperPac™.
Low Battery Indicator: Flashing low battery icon
on display.
Multiple Paired Contacts: Ensures quick, error-free
insertion and backup reliability.
Recharge Rate: 80% in 7.5 hours, 100% in 9.5 hours.
(For high capacity batteries)

MONITOR/DISPLAY
Input: 3-lead or 5-lead patient cable, paddles, or multipurpose hands-free adapter.
Size: 6.4 inch (16.5 cm) diagonal, non-fade.
Type: LCD Monochromatic.
Display Resolution: 640 x 480 pixels.
Sweep Speed: 25 mm / sec.
Lead Selections: Paddles (Pads), I, II, III, AVR, AVL, AVF, V.
Frequency Response: (User-selectable).
• 2 to 20 Hz Limited mode
• 0.5 to 40 Hz Monitor mode
• 0.05 to 150 Hz Diagnostic mode
(automatically sets chart recorder response)
Common Mode Rejection: Complies with AAMI EC13-1992 section 3.2.9.10.
Heart Rate Meter: 20 to 300 BPM.
Heart Rate Alarms: User-selectable.
Size: 0.125, 0.25, 0.5, 1, 2, 4 cm/mv and auto-ranging.

RECORDER
Type: High-resolution thermal array.
Annotation: Time, date, ECG lead, ECG gain, heart rate, defibrillation and pacing
parameters and treatment summary ACLS events.
Paper Width: 50 mm.
Paper Speed: 25 mm/sec.
Delay: 6 seconds.
Frequency Response: Automatically set to monitor's frequency response.
Treatment Summary: Records defibrillation, pacing and status events. Automatically
logs into memory the type of event, time and ECG sample.
Tx Summary Log: 28 ECG events or 300 non-ECG events.
Record Modes: Manual and automatic

pic30.jpg

If you would like to receive more information on any of the products featured on our website, please click here.

Software * Services * Technology for the Medical Community