GENERAL Weight: 13.5 lbs (6.1 kg) (basic configuration). Size:
16.7 x 12.5 x 5.3 inches; 41.75 x 32 x 13 cm. Operating: Time: 2 hours Temperature*: 0 to 45° C Humidity* (NC):
15 to 95% RH Vibration*: MIL-STD 810E Shock / Drop*: MIL-STD 810E Altitude: Up to 15,000 feet Transport and Storage:
Temperature (-30 to 70˚C), humidity (15 to 95% non-condensing), atmospheric pressure ( 860 to 1060 hPa), shock/vibration
(ISTA 1A) Enclosure Protection*: Solid Foreign Object: IEC 529, IP2X Water: IEC 529, IPX4. * All tests performed
per AAMI DF-2 Defibrillation Standard. Operating Power: Battery pack or auxiliary power supply.
DEFIBRILLATOR Waveform: Biphasic:
Low current, impedance compensated, optimized terminating current, truncated exponential biphasic. Energy Select: External:
2, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, 360 J. Internal: 2, 3, 4, 5, 6, 7, 8, 9, 10, 20, 30, 50 J. Charge Time:
< 7 seconds @ 360 J (with a charged battery). < 15 seconds @ 360 J (no battery, 90% AC mains voltage) Charge Indicator:
Audible and graphic. Output: Adult paddles, internal paddles, peds adapter and multipurpose hands-free adapter available. Synchronizer:
Delivers energy within 60 msec. Disarm: Front panel switch.
POWER Mains Voltage: 100-240VAC (auto ranging) Mains
Frequency: 50/60Hz
BATTERY - (User replaceable) High Capacity Type: NiMH 12 volt - Welch Allyn SuperPac™. Low
Battery Indicator: Flashing low battery icon on display. Multiple Paired Contacts: Ensures quick, error-free insertion
and backup reliability. Recharge Rate: 80% in 7.5 hours, 100% in 9.5 hours. (For high capacity batteries)
MONITOR/DISPLAY Input:
3-lead or 5-lead patient cable, paddles, or multipurpose hands-free adapter. Size: 6.4 inch (16.5 cm) diagonal, non-fade. Type:
LCD Monochromatic. Display Resolution: 640 x 480 pixels. Sweep Speed: 25 mm / sec. Lead Selections: Paddles (Pads),
I, II, III, AVR, AVL, AVF, V. Frequency Response: (User-selectable). • 2 to 20 Hz Limited mode • 0.5
to 40 Hz Monitor mode • 0.05 to 150 Hz Diagnostic mode (automatically sets chart recorder response) Common
Mode Rejection: Complies with AAMI EC13-1992 section 3.2.9.10. Heart Rate Meter: 20 to 300 BPM. Heart Rate Alarms: User-selectable. Size:
0.125, 0.25, 0.5, 1, 2, 4 cm/mv and auto-ranging.
RECORDER Type: High-resolution thermal array. Annotation:
Time, date, ECG lead, ECG gain, heart rate, defibrillation and pacing parameters and treatment summary ACLS events. Paper
Width: 50 mm. Paper Speed: 25 mm/sec. Delay: 6 seconds. Frequency Response: Automatically set to monitor's frequency
response. Treatment Summary: Records defibrillation, pacing and status events. Automatically logs into memory the type
of event, time and ECG sample. Tx Summary Log: 28 ECG events or 300 non-ECG events. Record Modes: Manual and automatic
|
 |
|