Technical Specifications

DEFIBRILLATOR
Output: Welch Allyn Orbital™ Biphasic Truncated exponential.
Energy Select: 200, 300, 360J or user configurable among the following energy options: 150, 200, 300, 360
Charge Time: Less than 5 seconds to 150J, less than 15 seconds to 360J.
Analysis Time: 4-16 sec.
Audible Prompts: 18 audible prompts.
Visual Prompts: 20 text screen prompts.
Controls: Four buttons - On/Off, Shock, and 2 software configurable buttons.
Accuracy: ±10% or 1J (whichever is greater) at 50Ω.

DISPLAY
Size: 54 x 36 mm.
Type: Backlit monochrome LCD
Display Resolution: 160 x 100 pixels.
Low Battery Indicator: Battery Icon gauge on display with 10 capacity levels.
Backlight: LED.

BATTERY
Non-Rechargeable: LiMnO2 12V, 2.6 Ah.
Capacity: 90 discharges at 360 Joules or 125 discharges at 200 Joules or 150
discharges at 150 Joules or 5 hours ECG monitoring.
Shelf Life: (25°C±15°C) 10 years (5 years storage + 5 years standby); 5 years standby
(after installation).*
* Capacity may be diminished at operating temperature extremes.

GENERAL
Weight: 3.5 lbs. with battery (1.6 kg)
Size: 8.25" x 6.875" x 2.75" (210 x 175 x 70 mm)
Operating Temperature: 0° to 50°C
Storage Temperature: -30° to 70°C (without battery)
Humidity: Up to 95% or 10° to 95°C (non-condensing)
Altitude: Up to 15,000 feet (-150 to 4570 m)
Shock: Mil Std 810E method 516.4, procedure 1 (40G, 6-9ms pulse, 1/2 sine each axis)
Vibration: Mil Std 810 method 514.4, category 1
Water resistance: IEC 529 IPX4
Drop Test: Mil Std 810F method 516.5 Procedure IV (1 meter drop onto any edge,
corner, or surface)

ELECTRICAL ISOLATION
Power: Unit operates on internal battery only.
External Electrical Connections: No external devices are attached to the unit.
Risk Current Category: Class II equipment with defibrillator-proof BF type patient
applied parts (as per definition of IEC 60601-1 standard.)

ELECTROMAGNETIC COMPATIBILTY
Radiated Emissions: EN55011; CISPR11 B
ESD: EN61000-4-2; 8KV air 6KV contact
Radiated Susceptibility: EN61000-4-3 (10 V/M (20 v/m EN60601-2-4))
Magnetic Field Emissions: MIL STD RE101 (AAMI DF 24.3.18.1)
Magnetic Field Susceptibility: MIL STD RS101 1 Gauss, 47 Hz-1.8 KHz.

DATA MANAGEMENT
Event Documentation: Internal.
Internal Memory Capacity: 1MB: 250 4-sec ECG samples or 3000 time stamped events.
Quick Report: Treatment Summary, Event Log, Test Log.
Communication: RS-232 via Infrared (IR) port to PC or printer.